Choosing the Correct Test Target

Test targets are useful when evaluating or calibrating an imaging system's performance or image quality. This could include troubleshooting the system, certifying or evaluating measurements, as well as establishing a foundation to ensure the system works well with another. Image quality can be defined by different components, particularly resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion; therefore, one or more types of test targets may be necessary or helpful depending upon the type of system being constructed or what needs to be measured. Fortunay, an array of targets exists that cater towards specific systems including cameras, visual displays, or even a single, thin lens. To be able to choose the correct test target, it is important to first understand the components of image quality.

 

COMPONENTS OF IMAGE QUALITY

Resolution

Resolution is an imaging system's ability to distinguish object detail. It is often expressed in terms of line-pairs per millimeter (lp/mm) as seen in Figure 1. A low resolution image usually lacks fine detail and is often blurry, whereas a high resolution image is highly detailed and clear.

Figure 1: Relation of Line-Pairs to Square Waves

 

To illustrate this concept, imagine two squares being imaged onto pixels of a CCD camera. Assuming that the primary magnification of the lens is such that one square fills one camera pixel (Figure 2a), if there is no space between the filled pixels, it will appear as one large, red rectangle. However, if "white space", or space distinctively different from the original pixel color, is found between the pixels, the camera will be able to distinguish among the two squares (Figure 2b). Hence, the pairing of the red square and "white space" becomes one lp/mm, which corresponds to two separate pixels.

Figure 2: Pair of Red Squares Unresolved (a) vs. Resolved (b)

 

Contrast

Contrast is a measurement of the separation between the light and dark regions of an image. More specifically, contrast is a change in the intensity or brightness from one point to another. It affects how effectively the differences between the object and the shades of gray in the background are reproduced. An image with the highest contrast is one in which black is truly black and white is truly white, without any shades of gray in between. As contrast is reduced, the distinction between black and white begins to blur, in a very literal sense, and shades of gray appear (Figure 3).

Figure 3: Contrast in Relation to Pixels

Contrast is often expressed in terms of percentage (%) and is calculated by using Maximum Intensity (Imax) and Minimum Intensity (Imin), as expressed in Equation 1. It can also be represented by a periodic function (i.e. square wave or sine wave), or a function that alternates regularly and instantaneously between two levels.

 

Modulation Transfer Function (MTF)

Modulation Transfer Function, or MTF, is a measurement of an imaging lens' ability to transfer contrast from the object plane to the image plane at a specific resolution. The object and image planes are the spatial areas where the object and the image preside. The object plane is in front of the imaging system, and the image plane is either in front or behind the imaging system depending on whether the image is real or virtual. MTF is expressed with respect to image resolution (lp/mm) and contrast (%), as seen in Figure 4. Typically, as resolution increases, contrast decreases until a cut-off point, at which the image becomes irresolvable and grey.

Figure 4: Example MTF Curve of 0.13X PMAG Imaging Lens

 

Another component of MTF, in addition to the aforementioned resolution and contrast, is diffraction limit. Diffraction limit is a physical limit restricting a lens from being able to image points or edges perfectly. Since it is constrained by the wave nature of light, even a "perfectly" designed and manufactured lens cannot achieve diffraction limited performance. However, designers utilize a variety of methods to reduce aberrations and increase overall system accuracy in order to come as close as possible to reaching a system's ideal diffraction limit.

Correspondingly, a len's geometry contributes to its ability to reproduce good quality image. Lens Diameter (D), Focal Length (f) and f/# (Equation 2) all affect MTF.

f/# is the light gathering ability of a lens. As Lens Diameter increases, f/# decreases. Low f/# lenses collect the most light, thereby making them ideal for light restrictive applications. Although high f/# can improve an imaging lens' performance, increasing it too much can be detrimental because it can cause the diffraction limit to become progressively worse.

 

Depth of Field (DOF)

Depth of Field, or DOF, is the ability of a lens to maintain a desired amount of image quality as the object being viewed moves in and out of focus. DOF is defined with an associated resolution and contrast, as both suffer when an object is placed closer or farther from the optimal working distance (Figure 5). DOF also applies to objects with depth, since high DOF lenses can image the whole object clearly. DOF is estimated by a single value calculated from the diffraction limit, making it a theoretical approximation. It is difficult, however, to make a genuine comparison because many imaging lenses are not diffraction limited. For example, two lenses may have the same f/# (i.e. equal diffraction limit), but do not necessarily have similar performance or comparable DOF. Therefore, the only way to truly determine DOF is to use a test target to test it.

Figure 5: Geometric Representation of DOF for Low and High f/# Imaging Lenses

 

Distortion

Distortion is a type of geometrical aberration that causes a difference in magnification of the object at different points in the image. As light rays carry the image of an object through a system, various points are misplaced relative to the center of the field, or the central point of the image. Therefore, distortion is not an aberration that causes blurs, but an aberration that causes a dislocation. Distortion is calculated by:

 

where AD is the Actual Distance and PD is the Paraxial, or Predicted, Distance.

Distortion, represented by a percentage, can be either positive or negative. A positive percentage represents "pincushion" distortion, whereas a negative percentage represents "barrel" distortion. Figure 6 illustrates barrel and pincushion distortion compared to an ideal, perfectly square non-distorted image.

Figure 6: Barrel and Pincushion Distortion

 

Although distortion is present in almost all lenses, it can be corrected by using short focal lengths. Unfortunay, short focal length systems tend to suffer from more diffraction effects than their longer focal length counterparts. Correcting for one component of image quality undeniably affects another, a fact that must always be taken into account. An important concept to keep in mind is that distortion causes information about the object to be misplaced but not lost. The original, un-distorted image can be reconstructed through image analysis software.

 

TYPES OF TEST TARGETS

A variety of test targets exist to help characterize resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion in an imaging system. Use the following guide of the most popular targets to make choosing the correct test target as easy as possible.

 

1951 USAF Resolution Targets

Consist of horizontal and vertical bars organized in groups and elements. Each group is comprised of up to nine elements within a range of twelve groups. Every element is composed of three horizontal and three vertical bars equally spaced with one another within a group and corresponds to an associated resolution based on bar width and space. The vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution. These targets are very popular when considering a target for testing resolution.

 

Typical Applications

Testing Resolution in Applications such as Optical Test Equipment, Microscopes, High Magnification Video Lenses, Fluorescence and Confocal Microscopy, Photolithography, and Nanotechnology

 

IEEE Resolution Targets

Designed to characterize the amount of resolution a camera or display system is able to reproduce from an original image. Because resolution can be different throughout the field of view, both horizontal and vertical resolution can be measured in the center of the target as well as the four corners. IEEE Resolution Targets can also be used to check scanning, linearity, aspect, shading, and interlacing, as well as measure TV lines.

 

Typical Applications

Testing of Analog Imaging Systems

 

Ronchi Rulings

Consist of a square wave optic with constant bar and space patterns that carry a high contrast ratio. They are ideal for reticle and field calibration requirements, and often used for evaluating resolution, field distortion and parafocal stability. Ronchi Rulings are not limited to only calculating resolution; they can be used for diffraction testing.

 

Typical Applications

Testing the Parameters of Resolution and Contrast, Diffraction Testing

 

Distortion Targets

Used for calibrating imaging systems for distortion, which is a geometrical aberration that may misplace certain parts of the image. These targets consist of a grid of dots that are separated by various distances depending on the application.

 

Typical Applications

Lower Focal Length Lenses, Systems that Carry a Wide Field of View

 

Depths of Field Targets

Depths of Field Targets Directly test the depth of field in imaging systems without the use of calculations. The target should be mounted 45° from the face of the lens that is parallel to the object to be viewed; the scale of the target consists of horizontal and vertical lines that measure frequency in line pairs per mm (lp/mm).

 

Typical Applications

Circuit Board Inspection, Security Cameras

 

Star Targets

Star Targets Ideal for identifying focus errors, astigmatism, as well as other focus differing aberrations. The target consists of a circle formed with alternating black and white radial lines emanating from a central point. Because the lines taper, a continuous change in resolution is present and can be measured in both vertical and horizontal directions without repositioning.

 

Typical Applications

Alignment of a System, Assistance with Assembly, Comparing Highly Resolved or Magnified Imaging Systems

 

EIA Grayscale Targets

EIA Grayscale Targets Useful for testing optical and video inspection systems, and consist of a standard pattern and carry two scales, one linear and the other logarithmic, which is useful depending on the linearity of the detector being used. Each scale has nine steps that are acuy tuned for a precise halftone pattern.

 

Typical Applications

Optical and Video Inspection Systems, Evaluating Contrast Levels in Cameras

 

Color Checker Targets

Color Checker Targets Used to determine true color balance or optical density of any color rendition system. They may be expanded to include more squares with a different assortment of colors and act as a reference for testing and standardizing color inspection and analysis systems.

 

Typical Applications

Color Rendition Systems, Digital Cameras and Photography

 

Exploring 1951 USAF Resolution Targets

1951 USAF Resolution Targets have been and are currently a standard when considering a target that tests the resolution of an imaging system. They consist of horizontal and vertical bars organized in groups and elements. Each group is composed of six elements, and each element is composed of three horizontal and three vertical bars equally spaced with one another. There can be a total of twelve groups, with larger numbers used for higher resolution. For example, a standard resolution 1951 target consists of group numbers from -2 to 7, whereas a high resolution of -2 to 9; the element number is the same. The resolution is based on bar width and space, where the length of the bars is equal to five times the width of a bar (Figure 7). One Line Pair (lp) is equivalent to one black bar and one white bar. Vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution.

Figure 7: 1951 USAF Target Specifications

 

Qualitatively, the resolution of an imaging system is defined as the group and element combination directly before the black and white bars begin to blur together. Quantitatively, resolution (in terms of line pairs per millimeter, or lp/mm) can be calculated by:

 

It is important to keep in mind that calculating resolution with a 1951 USAF Target is subjective. In other words, it depends on who is looking at the target. Someone with 20/20 vision (using the Snellen Ratio) is able to discern higher resolution than someone with, for example, 20/25 or 20/30 vision. Even though the actual test yields precise resolution values, the user's vision can lead to imprecise measurements.

 

APPLICATION EXAMPLES

Example 1: Calculating Resolution with a 1951 USAF Resolution Target

When given a specified group and element number, one can easily calculate the resolution in lp/mm using Equation 4. For instance, if the vertical or horizontal bars start to blur at group 4 element 3, the resolution of the system can be designated as group 4 element 2. To quickly calculate resolution, use our 1951 USAF Resolution EO Tech Tool.

To convert lp/mm to microns (μm), simply take the reciprocal of the lp/mm resolution value and multiply by 1000.

 

Example 2: Understanding f/#

To understand the relationship between f/#, depth of field, and resolution, consider an example with a 35mm Double Gauss Imaging Lens (Figure 8). In this example, the lens will be integrated into a system which requires a minimum of 5 lp/mm (200μm) object resolution at 20% contrast. The diffraction limit , or cutoff frequency, is determined by Equation 7:

Figure 8: Graphical Representation of Resolution vs. f/# (Left) and DOF vs. f/# (Right) for a 35mm Double Gauss Imaging Lens

 

where λ is the wavelength of the system. For simplicity, Equation 7 assumes a non-aberrated, ideal system. Because this system is expected to have aberrations, though, the diffraction limit decreases with increasing f/#. Determining an ideal f/# for this system leads to calculating the highest possible depth of field. Comparing resolution vs. f/#, it is evident that below f/3, the lens is limited by aberrations and cannot obtain the minimum desired resolution. However, "sping down" or closing the iris reduces aberrations and improves DOF. At f/4.2, diffraction effects caused by the optical elements within the imaging lens become more prominent than the effects from aberrations; this is the point at which the lens becomes diffraction limited. Beyond f/4.2, closing the aperture increases DOF, but reduces resolution. At f/13.5, the diffraction limit defines the extent of the desired resolution. Beyond f/13.5, resolution continues to decrease while DOF continues to increase. In this particular example, f/13.5 is the ideal f/# for an optimum depth of field at a minimum resolution.

聯系我們

地址:江蘇省江陰市人民東路1091號1017室 傳真:0510-68836817 Email:sales@rympo.com
24小時在線客服,為您服務!

版權所有 © 2025 江陰韻翔光電技術有限公司 備案號:蘇ICP備16003332號-1 技術支持:化工儀器網 管理登陸 GoogleSitemap

在線咨詢
QQ客服
QQ:17041053
電話咨詢
0510-68836815
關注微信
日韩欧美国产一区二区_国产精品国模大尺度视频_在线中文一区_久久综合九色99
欧美精品videossex性护士| 国产精品视频内| 一区二区福利| 一区二区三区四区五区精品视频| 国产精品午夜久久| 国产在线视频不卡二| 影视先锋久久| 亚洲素人在线| 美女任你摸久久| 国产精品视频福利| 亚洲人成网站777色婷婷| 亚洲无线一线二线三线区别av| 亚洲欧美影音先锋| 欧美激情一区二区三级高清视频| 免费在线观看精品| 国产精品毛片a∨一区二区三区| 国产亚洲一级高清| 一本久道久久久| 老司机精品久久| 国产精品视频网| 99视频精品免费观看| 久久噜噜噜精品国产亚洲综合 | 性色一区二区| 欧美电影专区| 一区二区亚洲精品| 午夜精品视频一区| 欧美日韩一区二区视频在线| 在线高清一区| 久久精品一区蜜桃臀影院| 国产精品久久久久久久久免费樱桃 | 91久久精品美女高潮| 欧美一级淫片aaaaaaa视频| 欧美激情在线观看| 亚洲国产精品综合| 久久综合久久综合这里只有精品 | 亚洲精品久久久久久久久| 久久久999| 国产综合久久久久久| 亚洲一区视频| 国产精品午夜在线观看| 午夜精品久久久久影视| 国产精品美女久久福利网站| 亚洲午夜高清视频| 欧美午夜国产| 亚洲尤物视频在线| 国产欧美 在线欧美| 亚洲欧美一级二级三级| 国产欧美精品va在线观看| 欧美中文字幕视频| 在线精品亚洲| 欧美成人免费全部| 亚洲精品一区二区三区福利| 欧美片第1页综合| 亚洲图片欧洲图片av| 国产精品va在线播放| 亚洲午夜免费福利视频| 国产欧美一区二区三区在线看蜜臀| 亚洲网站在线播放| 国产精品一区二区在线观看网站| 性做久久久久久免费观看欧美| 国产视频一区三区| 免费一区视频| 羞羞视频在线观看欧美| 在线免费观看日本欧美| 欧美日韩亚洲一区| 久久激情五月激情| 一本色道久久综合一区| 国产一区二区久久精品| 欧美巨乳在线| 久久九九热re6这里有精品| 亚洲精品精选| 国产亚洲福利| 欧美成人首页| 亚洲午夜激情在线| 国产综合av| 欧美日韩在线大尺度| 久久精品日韩欧美| 一区二区三区四区在线| 国产日韩欧美91| 欧美日韩亚洲一区三区 | 国产精品高清在线| 久久久水蜜桃| 亚洲深夜福利在线| 在线成人h网| 国产农村妇女精品一区二区| 欧美1区2区| 亚洲欧美日韩成人| 99成人在线| 91久久亚洲| 在线观看日韩av电影| 欧美偷拍另类| 欧美精品免费看| 久久―日本道色综合久久| 亚洲免费视频一区二区| 日韩亚洲精品视频| 亚洲国产美女精品久久久久∴| 国产日韩精品视频一区| 国产精品每日更新| 欧美精品久久天天躁| 老色鬼久久亚洲一区二区| 久久国内精品自在自线400部| 国产精品99久久久久久久女警| 亚洲第一区在线| 黄色精品一区| 国产一区二区三区黄| 欧美视频观看一区| 欧美人在线观看| 欧美国产精品v| 欧美高清免费| 欧美区国产区| 欧美视频在线观看一区二区| 欧美区高清在线| 欧美三级午夜理伦三级中视频| 欧美电影免费观看高清| 猫咪成人在线观看| 美国成人直播| 欧美激情一区二区| 欧美日韩久久精品| 欧美日韩亚洲不卡| 国产精品毛片在线看| 国产精品永久免费视频| 国产欧美一区二区三区另类精品| 国产精品视频一区二区高潮| 国产精品区一区二区三区| 国产老肥熟一区二区三区| 国产亚洲午夜| 亚洲国产精品一区在线观看不卡 | 欧美日本国产| 欧美午夜欧美| 国产欧美一级| 亚洲电影下载| 亚洲视频欧美视频| 久久精品欧美日韩精品| 欧美成人免费全部| 欧美日韩一区二区欧美激情| 国产精品xxx在线观看www| 国产日韩精品综合网站| 亚洲黄一区二区三区| 亚洲一二三四区| 久久一区中文字幕| 欧美日韩亚洲精品内裤| 国内欧美视频一区二区| aa日韩免费精品视频一| 久久久久久尹人网香蕉| 欧美精品入口| 国内成人精品2018免费看| 亚洲三级免费| 久久久久国产精品午夜一区| 欧美日韩精品中文字幕| 国产视频亚洲| 一区二区精品在线观看| 久久一本综合频道| 国产精品人人爽人人做我的可爱| 亚洲高清在线观看| 欧美亚洲在线观看| 欧美日韩免费观看一区| 精品999网站| 亚洲尤物影院| 欧美日本韩国在线| 亚洲激情国产| 久久久夜夜夜| 国产亚洲二区| 亚洲欧美一级二级三级| 欧美精品97| 亚洲第一区在线| 久久久亚洲高清| 国产亚洲激情| 欧美一级专区| 国产伦精品一区二区三区高清| 一本久久综合亚洲鲁鲁五月天| 蜜桃精品久久久久久久免费影院| 国产欧美日韩精品一区| 正在播放日韩| 欧美亚一区二区| 在线亚洲精品| 国产精品久久久久久久久久ktv| 亚洲精品五月天| 欧美二区在线播放| 亚洲第一福利社区| 欧美成人免费视频| 亚洲区一区二| 欧美日韩视频在线| 亚洲视频免费在线观看| 国产精品99免视看9| 在线一区视频| 国产精品视频网站| 欧美一区观看| 狠狠色伊人亚洲综合网站色| 欧美一区二区三区日韩| 国产亚洲欧美日韩精品| 久久国内精品视频| 伊人蜜桃色噜噜激情综合| 久久久久久高潮国产精品视| 国产午夜精品一区二区三区视频 | 国产精自产拍久久久久久| 中文亚洲免费| 国产精品永久免费在线| 久久久久久综合| 亚洲精品日韩综合观看成人91| 欧美激情第1页|